Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00886-1
Product Name/Description Spectra Optia Apheresis System - Return Line Air Detector (RLAD)

ARTG Number: 130529
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 27/08/2014
Responsible Entity Terumo BCT Australia Pty Ltd
Reason/Issue Terumo BCT has redesigned the Return Line Air Detector (RLAD) component of the Spectra Optia Apheresis System. This redesign will further reduce the low occurrence of false air detections during a procedure; the plan has been to install the replacement component on devices in which a RLAD false air detect has occurred.
Recall Action Recall for Product Correction
Recall Action Instructions Terumo BCT is replacing the existing RLAD sensor with the redesigned component on all Spectra Optia systems in Australia. If false air detections have occurred on a system in the last 12 months the upgrade will be scheduled as soon as possible. Where no issues have been reported the RLAD sensor will be replaced during the next scheduled preventative maintenance visit. This action has been closed-out on 03/05/2016.
Contact Information 02 9429 3600 - Terumo BCT