Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00879-1 |
Product Name/Description |
LIAISON Biotrin Parvovirus B19 IgM
Part Number: 317010
Lot Number: 129022X, 129022X/1
Expiry Date: 12 December, 2014
ARTG Number: 203754
An in vitro diagnostic medical device (IVD) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
11/08/2014 |
Responsible Entity |
|
Reason/Issue |
DiaSorin has become aware that the kit lots identified have a potential to produce an increased rate of positive results when used on LIAISON XL analyzer. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
DiaSorin is recommending the retesting of all specimens resulting in an initial positive on the LIAISON XL analyzer unless, as part of the laboratory policy, these samples have already been confirmed by other method or by clinical evidence. The affected lots can still be used, but confirming all initially positive results with a different method or analyzer. DiaSorin is offering Parvovirus B19 IgM EIA, code V619IM free of charge to perform the retesting. This action has been closed-out on 03/06/2016. |
Contact Information |
1300 259 835 - DiaSorin Australia Pty Ltd |