Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00879-1
Product Name/Description LIAISON Biotrin Parvovirus B19 IgM

Part Number: 317010

Lot Number: 129022X, 129022X/1

Expiry Date: 12 December, 2014

ARTG Number: 203754

An in vitro diagnostic medical device (IVD)
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 11/08/2014
Responsible Entity DiaSorin Australia Pty Ltd
Reason/Issue DiaSorin has become aware that the kit lots identified have a potential to produce an increased rate of positive results when used on LIAISON XL analyzer.
Recall Action Recall for Product Correction
Recall Action Instructions DiaSorin is recommending the retesting of all specimens resulting in an initial positive on the LIAISON XL analyzer unless, as part of the laboratory policy, these samples have already been confirmed by other method or by clinical evidence. The affected lots can still be used, but confirming all initially positive results with a different method or analyzer. DiaSorin is offering Parvovirus B19 IgM EIA, code V619IM free of charge to perform the retesting. This action has been closed-out on 03/06/2016.
Contact Information 1300 259 835 - DiaSorin Australia Pty Ltd