Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00877-1
Product Name/Description Transpac Monitoring Kits

Arterial Safeset Transpac IT w/3 ml Reservoir and Single CSP, With Velcro Arm Strap, Patient Mount
Item Number: 011-46104-42
Lot Numbers: 2841944, 2841945, 2859326, 2865873 & 2869445

Arterial Transpac IT Monitoring Kit w/Safeset Reservoir, 03 ml Flush Device and 2 CSP
Item Number: 011-46106-72
Lot Numbers: 2841690, 2841691 & 2859331

ARTG Number: 206780
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 8/08/2014
Responsible Entity ICU Medical Australia Pty Ltd
Reason/Issue ICU Medical has become aware of an issue with a component where the connection on an arterial monitoring kit may crack and cause a fluid leak. A leaking luer may have the potential effect of blood leakage or air embolism which may be life threatening or cause permanent severe injury or illness to the patient. There have been no reports of patient injury due to this problem.
Recall Action Recall
Recall Action Instructions ICU Medical Australia Pty Ltd is requesting their customers to inspect their stocks and quarantine all units from the affected lots. This action has been closed-out on 03/06/2016.