Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00874-1 |
Product Name/Description |
Chemistry 1 Calibrator used on Dimension Vista Systems (Magnesium assay calibrator). An in vitro diagnostic medical device (IVD).
Catalogue Number: KC110
Siemens Material Number: 10445169
Lot Number: 3GM081
ARTG number: 182222 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
8/08/2014 |
Responsible Entity |
|
Reason/Issue |
Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista CHEM 1 CAL lot 3GM081. Investigation by Siemens confirms that QC and patient samples show low recovery of MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
CHEM 1 CAL lot 3GM081 expired on 1 July 2014; however, Siemens is advising their customers to immediately calibrate MG with an alternate lot of CHEM 1 CAL if a calibration of MG with this lot of CHEM 1 CAL is still active. Siemens recommends that the content of this letter and review of previous testing should be discussed with your Medical Director. This action has been closed-out on 03/06/2016. |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |