Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00874-1
Product Name/Description Chemistry 1 Calibrator used on Dimension Vista Systems (Magnesium assay calibrator). An in vitro diagnostic medical device (IVD).

Catalogue Number: KC110

Siemens Material Number: 10445169

Lot Number: 3GM081

ARTG number: 182222
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 8/08/2014
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens Healthcare Diagnostics has received complaints for low recovery of Magnesium (MG) Quality Control (QC) and patient samples following calibration of MG with the Dimension Vista CHEM 1 CAL lot 3GM081. Investigation by Siemens confirms that QC and patient samples show low recovery of MG by 0.25 mg/dL [0.10 mmol/L] across the assay range following calibration with 3GM081.
Recall Action Recall for Product Correction
Recall Action Instructions CHEM 1 CAL lot 3GM081 expired on 1 July 2014; however, Siemens is advising their customers to immediately calibrate MG with an alternate lot of CHEM 1 CAL if a calibration of MG with this lot of CHEM 1 CAL is still active. Siemens recommends that the content of this letter and review of previous testing should be discussed with your Medical Director. This action has been closed-out on 03/06/2016.
Contact Information 1800 310 300 - Siemens Technical Support Centre