Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00834-1
Product Name/Description Discovery XR656 Wall Stand System (X-ray system)

Part Numbers: 5374989 and 5397837

Serial numbers: 00000100365HL6, 00000099847HL6, 00000109154HL5, 00000108716HL2, 00000109561HL1, 00000109564HL5

ARTG Number: 167617
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 25/07/2014
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE Healthcare has recently become aware of two FlashPad detector falls due to a broken nylon support hook. A detector fall could result in an injury to a patient or operator. There have been reported incidents of detector falls, but no injuries have been reported.
Recall Action Recall for Product Correction
Recall Action Instructions Users are advised that if they observe either one of the nylon support hooks as broken or loose on the Discovery XR656 wall stand system, immediately stop use and contact their local GE Healthcare Service Representative. If the hooks are not broken, users may continue to use the wall stand system. GE Healthcare is correcting all affected systems by replacing the nylon hooks. This action has been closed-out on 03/06/2016. This action has been closed-out on 03/06/2016.