Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00820-1 |
Product Name/Description |
VITROS Immunodiagnostic Products CA 19-9 Calibrators and Reagent Packs An in vitro diagnostic medical device (IVD)
VITROS Immunodiagnostic Products CA 19-9 Calibrators Product Code: 6800035
VITROS Immunodiagnostic Products CA 19-9 Reagent Pack Product Code: 6800040
Lot Number: 1320
Expiry Date: 12 September 2014 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
24/07/2014 |
Responsible Entity |
|
Reason/Issue |
Ortho-Clinical Diagnostics (OCD) received complaints of results for patient and proficiency samples were elevated in some instances. Proficiency sample results when using VITROS CA 19-9 kit lot 1320, compared to previous kit lots used and analysed (OCD internal testing), found an overall average positive bias of approximately +30% for samples within the normal range (<=37 U/mL) and +17% for samples above the normal range (>37 U/mL). |
Recall Action |
Recall |
Recall Action Instructions |
Laboratories are being advised to immediately discontinue using and discard all remaining inventory of VITROS CA19-9 Calibrators and Reagent Packs, Lot 1320. Review previously reported results using VITROS CA 19-9 Reagent Packs, Lot 1320 and discuss any concerns regarding previously reported results with your Laboratory Medical Director or the requesting physician to determine the appropriate course of action. Ortho-Clinical Diagnostics (OCD) will provide credits/replace stock of any product affected by this recall. This action has been closed-out on 12/08/2016. |
Contact Information |
1800 032 359 - Ortho-Clinical Diagnostics Customer Technical Service |