Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00795-1
Product Name/Description TXR-ME (Tri-Ex Extraction Balloon with Multiple Sizing)

(275 cm Length only)

Serial Numbers: W3372236(1), W3376524(1), W3427509(2), W3406080(1), W3388883(2), W3403192(2), W3334395(1), W3374821(1), W3376524(1), W3308765(2), W3333058(2), W3415262(2)

ARTG number: 133685
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 18/07/2014
Responsible Entity William A Cook Australia Pty Ltd
Reason/Issue Cook Medical has received complaints that some Tri-Ex extraction balloons may not fit through endoscopes with a 2.8 mm channel. The extraction balloon is labelled for use with a 2.8 mm endoscope.
Recall Action Recall
Recall Action Instructions Cook medical is requesting their customers to review their stock for the affected lot numbers and quarantine any affected product in preparation for return. This action has been closed-out on 03/06/2016.
Contact Information 1800 777 222 - William A. Cook Australia Pty Ltd