Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00784-1
Product Name/Description Brainlab Patient Data Manager

Product Version Number: 2.0

Sub Versions: 2.0.0, 2.0.1 & 2.0.2

ARTG Number: 96517
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 14/07/2014
Responsible Entity Brainlab Australia Pty Ltd
Reason/Issue Brainlab has identified two potential problems when loading image data that consists of non-square pixels with PDM 2.0.

1- Display of image data: When loading non-square pixel image data with PDM 2.0, the data set is not correctly transferred and displayed in the corresponding Brainlab navigation or planning software.
2- Dose calculation: It has been determined that non-square pixel data is not suitable for dose calculation with the current Brainlab iPlan RT systems.
Recall Action Recall for Product Correction
Recall Action Instructions End users are advised to not load image data sets with non-square pixels if Brainlab Patient Data Manager version 2.0 (including subversions 2.0.0, 2.0.1 and 2.0.2) is installed at your hospital. Brainlab will provide a software update to affected customers to correct this issue. This action has been closed-out on 01/03/2016.