Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00757-1
Product Name/Description TRIGEN INTERTAN Intramedullary Nail 13MM X 20CM 125D

Batch Number: 14CM03676
Model Number: 71675206

ARTG Number: 104370
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 8/07/2014
Responsible Entity Smith & Nephew Surgical Pty Ltd
Reason/Issue The manufacturer has identified a manufacturing error which results in the proximal lag screw being positioned incorrectly. In the event an incorrectly positioned proximal lag screw hole inhibits drilling and causes the user to hit the nail, the following issues could occur: the nail is removed and/or the procedure is delayed, the nail’s fatigue life is reduced, causing the nail to fail within a short period of time.
Recall Action Recall
Recall Action Instructions The affected customer has been requested to inspect their inventory and quarantine the affected stock. The local Smith & Nephew Surgical representative will contact the customer to arrange for the product to be returned.
Contact Information 02 9857 3999 - Smith & Nephew Surgical