Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00736-1
Product Name/Description Slide-in Plug Adaptor for Charger used with Haag Streit UK Perkins MkIII Tonometer

Tonometer Product Code: CC-5806004

Serial Numbers: 274, 275, 276, 731, 732, 733, 453, 490, 452, 360

ARTG Number: 151673
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 4/07/2014
Responsible Entity Device Technologies Australia Pty Ltd
Reason/Issue Haag Streit has identified a potential issue with the slide-in adaptor plug for the Perkins MkIII Tonometer charger. An incident was reported to Haag-Streit where the use of the Adaptor tripped a customer’s circuit breaker. Upon investigation it was found the adaptor contained an electrical fault. This electrical fault has the potential to cause arcing within the adaptor, which poses a potential fire hazard. The risk when using the adaptor is most apparent if the adaptor is used in an un-fused circuit. The risk of harm and damage can be reduced if the unit is plugged into a conforming power outlet with a circuit breaker.
Recall Action Recall for Product Correction
Recall Action Instructions Users are advised to isolate the affected Plug Adaptors from use. A DTA representative will contact users in regards to replacement of affected stock.