Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00736-1 |
Product Name/Description |
Slide-in Plug Adaptor for Charger used with Haag Streit UK Perkins MkIII Tonometer
Tonometer Product Code: CC-5806004
Serial Numbers: 274, 275, 276, 731, 732, 733, 453, 490, 452, 360
ARTG Number: 151673 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
4/07/2014 |
Responsible Entity |
|
Reason/Issue |
Haag Streit has identified a potential issue with the slide-in adaptor plug for the Perkins MkIII Tonometer charger. An incident was reported to Haag-Streit where the use of the Adaptor tripped a customer’s circuit breaker. Upon investigation it was found the adaptor contained an electrical fault. This electrical fault has the potential to cause arcing within the adaptor, which poses a potential fire hazard. The risk when using the adaptor is most apparent if the adaptor is used in an un-fused circuit. The risk of harm and damage can be reduced if the unit is plugged into a conforming power outlet with a circuit breaker. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Users are advised to isolate the affected Plug Adaptors from use. A DTA representative will contact users in regards to replacement of affected stock. |
Contact Information |
1800 429 551 - Device Technologies Australia Pty Ltd |