TGA Recall Reference RC-2014-RN-00733-1
Product Name/Description AVS ARIA Implant Inserter

Item number: 48758500 (Located in Instrument Tray 48759040-T)

Lots numbers affected: 107400, 10G906, 110214

ARTG number: 178870
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/07/2014
Responsible Entity Stryker Australia Pty Ltd
Reason/Issue Stryker has received reports of fracture of the vertebral PEEK spacers when implanted with the AVS ARIA Implant Inserter. According to the reported complaints, the proximal end of the implant either cracks or the tip breaks off in one fragment. There have been no reports of the broken fragment remaining in the surgical site.
Recall Action Recall
Recall Action Instructions Stryker is advising their customers to quarantine any affected lots and will arrange for the return of any affected lots from the field.
Contact Information 1800 803 601 - Stryker Australia