Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00716-1
Product Name/Description Pasteur Pipettes for use in IVF

Affected product codes:
PP-5.75-90, Box of 90, 30 pouches of 3 pipettes each , 5.75” (14.6cm) long
PP-9-90, Box of 90, 30 pouches of 3 pipettes each, 9” (22.9cm) long
PP-5.75-90PL, Box of 90, cotton plugged, 30 pouches of 3 pipettes each, 5.75” (14.6cm) long
PP-9-90PL, box of 90, cotton plugged, 30 pouches of 3 pipettes each, 9” (22.9cm) long

ARTG Number: 221911
Recall Action Level Retail
Recall Action Classification Class II
Recall Action Commencement Date 11/07/2014
Responsible Entity Origio Australasia Pty Ltd
Reason/Issue ORIGIO has identified a 'pinhole' packaging breach on the 3-pack pouched version of the Pasteur pipets. As a result, the sterility of the package cannot be guaranteed.
Recall Action Recall
Recall Action Instructions Origio is requesting end users to return the affected lots of pastuer pipettes to their distributor for replacement or credit.