Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00698-1
Product Name/Description NexGen Complete Knee Solution MIS Total Knee Procedure Stemmed Tibial Component, Precoat

Various sizes

ARTG Number: 96761
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/07/2014
Responsible Entity Zimmer Pty Ltd
Reason/Issue Zimmer is initiating a Hazard Alert for specific lots already implanted due to the potential that the threads may be out of specification. Zimmer received twelve complaints over a period of two and a half years alleging difficulty threading an MIS Drop Down Stem Extension or a Drop Down Stem Plug into the MIS Stemmed Tibial Component. There were no complaints received from Australia. The potential risks for this issue are:

· Additional bone loss may occur if the MIS tibial is cemented in place when the issue is identified and the surgeon elects to remove and replace.
· Early revision may occur due to loosening (if the MIS Stemmed Tibial Component is implanted without a drop down stem extension), and/or infection.
· Disassociation of the implanted articular surface may occur if the stem extension does not fully seat, thus causing the articular surface to not properly lock.
Recall Action Hazard Alert
Recall Action Instructions Zimmer is providing surgeons with details of the potential risks associated with out of specification thread. For patients implanted with a potentially affected device, Zimmer recommends continued normal post-operative follow-up. For more details, please see http://www.tga.gov.au/safety/alerts-device-nexgen-complete-knee-solution-140709.htm .
Contact Information 1800 803 263 - Zimmer Customer Service