Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00674-1
Product Name/Description DX-D 100 Digital Radiography X-Ray System

All Units

ARTG Number: 178856
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 25/06/2014
Responsible Entity AGFA HealthCare Australia
Reason/Issue When liquid comes in contact with the DX-D 100 touch screen, the device may incorrectly recognise this as user input altering device settings. Capacitive touch screens rely on the electrical properties of the human body to enable users to operate the device with very light touches of a finger. However, when liquid comes in contact with the touch screen it may incorrectly recognise this as user input. For example, it could unexpectedly change the exposure settings.
Recall Action Recall for Product Correction
Recall Action Instructions Users are advised that they must ensure their hands are dry before using the touch screen of the DX-D 100 because liquid residue may activate the action buttons on the touch screen. The following advice is provided in the customer letter in addition to referring users to the warning in the user manual::
- Do not operate touch screen monitor with wet hands.
- Do not let liquids come in contact with the touch screen while the DXD 100 is powered on.
- Always double check your parameter settings prior to exposing the patient.
Contact Information 1300 364 612 - Agfa HealthCare Service Help Desk