TGA Recall Reference RC-2014-RN-00653-1
Product Name/Description Puritan Bennett 840 Ventilator with 9.4 inch graphical user interface (GUI) assembly

ARTG Number: 182309
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 18/06/2014
Responsible Entity Covidien Pty Ltd
Reason/Issue Covidien has received customer reports that have included the GUI display screens (both upper and lower) becoming blank during use. This event interferes with the clinician’s ability to view and/or change ventilator setting or clinician-set patient alarm parameters. The ventilator continues to provide respiratory support during these events.
Recall Action Recall for Product Correction
Recall Action Instructions Covidien is advising that if users encounter a loss of display, they are advised to verify the patient’s respiratory and physiological stability ie, confirm the patient is receiving ventilatory support by observation of the expansion and contraction of the patient’s chest and assess current patient status by review of other monitoring (e.g., oxygen saturation, heart rate, blood pressure, etc). The patient should be transferred to an alternative ventilator.
Covidien will be replacing affected backlight inverter PCBAs and performing 10,000 hour preventative maintenance. This action has been closed-out on 29/01/2016.
Contact Information 1800 350 702 - Covidien Technical Services