Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00631-1 |
Product Name/Description |
Medtronic Duet External Drainage and Monitoring System (Cerebrospinal fluid drainage system)
Model Numbers: 46913 & 46914
Multiple lot numbers affected
ARTG Number: 162723 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
13/06/2014 |
Responsible Entity |
|
Reason/Issue |
The patient line tubing in the affected devices may become disconnected from the patient line stopcock. Investigations undertaken by Medtronic determined that this issue was caused by a combination of using tubing that exceeded the recommended shelf life as well as the cleaning process of the tubing during assembly. Disconnection of the patient line is more likely to occur during handling of the system by a healthcare professional. |
Recall Action |
Recall |
Recall Action Instructions |
Medtronic is advising their customers to cease use of the affected product and return any unused units. If any of the potentially affected products are in use it should be verified that the connections are secure and leak free. |
Contact Information |
02 9857 9234 - Medtronic Regulatory Affairs |