Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00629-1
Product Name/Description EmbryoScope Time-Lapse Incubator

Reference Numbers: ESD20000, ESD200000-EU, ESD200000-JAP
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 16/07/2014
Responsible Entity Gytech Pty Ltd
Reason/Issue When cleaning or disinfecting parts of the EmbryoScope time-lapse incubator, the metal clips which are attached to the embryo chamber and support the transparent plate positioned over the culture dish holder may come loose.
Recall Action Recall for Product Correction
Recall Action Instructions FertiiTech has updated the User Manual to include information on checking that the metal clips are in the correct position after cleaning. This amendment has been notified to end users.

This action was undertaken prior to notifying the Therapeutic Goods Administration.
Contact Information 03 9822 5911 - Gytech Pty Ltd