Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00627-1
Product Name/Description BD FACSCanto 10 colour configuration (Cytometer). An in vitro diagnostic medical device (IVD)

Serial Number: V65733800038

ARTG Number: 188752
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 18/06/2014
Responsible Entity Becton Dickinson Pty Ltd
Reason/Issue BD has determined that there is a problem with the laser cooling fans and the fluidics cart internal fan impacting BD FACSCanto. The instrument laser cooling fans and the internal fan of the associated fluidics cart are non-operational potentially affecting the life os the laser of the instrument. Both problems are caused by a reversed wiring configuration that occurred during manufacturing.
Recall Action Recall for Product Correction
Recall Action Instructions A BD service representative will contact the customer to arrange for correction. This action has been closed-out on 18/02/2016.
Contact Information 02 8875 7165 - BD National Service Manager