Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00620-1
Product Name/Description Single-width Airway Module E-miniC and Extension Modules N-FC & N-FCREC

All units serviced with Field Replaceable Unit catalogue number M1013204 between February 2012 and May 2014.

ARTG Number: 166230
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 6/06/2014
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue GE Healthcare has become aware of a potential safety issue due to failure of the carbon dioxide detector associated with Single-width Airway and Extension Modules. Failure of the carbon dioxide detector in Single-width Airway and Extension modules may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value may impair clinical decision making for both mechanically and spontaneously ventilated patients because of incorrect low CO2 values.

This action has been undertaken further to the previous recall action (RC-2014-RN-00543-1) since GE has identified that all units serviced with Field Replaceable Unit catalogue number M1013204 between February 2012 and May 2014 are affected by this issue.
Recall Action Recall for Product Correction
Recall Action Instructions GE is providing calibration instructions to assist their customers to determine if their module is affected. If the carbon dioxide reading is out of range after the first successful calibration procedure end users are advised to discontinue use and contact a GE service representative. This action has been closed-out on 02/02/2016.
Contact Information 1800 659 465 - GE National Call Centre