Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00619-1
Product Name/Description Alere Triage BNP Calibrator for Beckman Coulter

Catalogue Number: BS98202

Manufacturer Part Number: 98202

Lot Number: 328163

Expiration Date: 30/09/2014

ARTG Number: 215842
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 6/06/2014
Responsible Entity Inverness Medical Innovations Australia Pty Ltd T/A Alere
Reason/Issue The bottle labelled as Calibrator Level 5 actually contains Calibrator Level 4. This results in a calibration failure when attempting to calibrate using a calibrator kit which contains the affected Level 5 Calibrator.
Recall Action Recall
Recall Action Instructions Alere is requesting their customers to discontinue use of the affected lot and discard any remaining stock. Replacements will be provided for any unused stock.
Contact Information 07 3363 7711 - Alere Technical Support