Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00613-1
Product Name/Description Multi Lumen and Pressure Injectable Central Venous Catheters

Affected Lot Numbers: 71F14A0329, RF1082062, 71F13H0637, ZF3063390, 71F14C0868, ZF1124891, ZF3051182, 71F13K0810, 71F13J1589, 71F14C0173, ZF2052625, 71F13H0972, 71F13K1389, 71F13K1796, 71F13K0429, ZF2111681, 71F13H1291, 71F13L0166, 71F14A0120, 71F13K1808, 71F14A1404, 71F13K0325, 71F13L1212, 71F14B0386, ZF3052590, RF0127942, 71F13K1625, ZF2123457, 71F14B0029, RF1069981, RF0087867, ZF3040526, 71F14B0377

ARTG Number: 142543
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 18/06/2014
Responsible Entity Mayo Healthcare Pty Ltd
Reason/Issue There is a low level of occurrence (<0.02%) of the lumen exit and lumen size on the extension line being misprinted on central venous catheters (CVC). As an example, the distal lumen may be printed with the word 'medial 18 GA' instead of 'distal 16GA'. Misprinted labelling can cause user dissatisfaction with flow rates (ml/hr) in the case the lumen would be expected to perform as labelled for high volume gravity infused (the mismarked 16GA Lumen) since in reality this is an 18GA Lumen and gravity flow is approximately half (1/2) the expected flow rate in ml/hr for a 16GA lumen. This could result in a worst case scenario clinical intervention to use a pressure bag or pump to achieve adequate flow rates. The colour coding of all hubs is correct and corresponds to the Arrow system.
Recall Action Recall for Product Correction
Recall Action Instructions Mayo Healthcare Pty Ltd is advising their customers to be aware of the incorrect labelling.
Contact Information 1300 360 226 - Mayo Healthcare National Customer Service