Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00608-1 |
Product Name/Description |
ADVIA Centaur, ADVIA Centaur XP and ADVIA Centaur CP - Troponin Ultra ReadyPack. An in vitro diagnostic medical device (IVD).
Catalogue Numbers: 02789602 (100 tests) and 02790309 (500 tests)
Lot Numbers:
Kit Lots ending in 078 Expiration: 18 July, 2014
Kit Lots ending in 079 Expiration: 18 July 2014
Kit Lots ending in 082 Expiration: 27 September, 2014
Kit Lots ending in 083 Expiration: 29 November, 2014
Kit Lots ending in 084 Expiration: 28 December 2014 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
5/06/2014 |
Responsible Entity |
|
Reason/Issue |
Manufacturer confirms that the solid phase reagent in some of the ADVIA Centaur TNI-Ultra ReadyPacks, lot 084, is darker in appearance, creating a potential for incorrect control and patient results. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is providing work around instructions to users. Only one TNI-Ultra ReadyPack is to be used on a system at one time. Patient results produced by the assay are acceptable if they follow acceptable calibration with valid quality control results. Kit lots ending in 086 or higher do not require calibration of each ReadyPack, refer IFU for the calibration interval. This action has been closed-out on 11/04/2016. |
Contact Information |
1800 310 300 - Siemens Technical Centre |