Product Name/Description RAPIDPoint 500 Measurement Cartridges. An in vitro diagnostic medical device (IVD)

Part numbers: 10844813, 10491447, 10491448 & 10491449

Serial Numbers: 2334500011 to 2412705915

ARTG number: 176642
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 3/06/2014
Responsible Entity Siemens Ltd Diagnostics Division
Reason/Issue Siemens Healthcare Diagnostics has received several reports from customers of one or more electrolyte parameters (e.g., Na+, Ca++, K+, Cl-) reporting a “-----?” error code when AQC, QC, or CVM is run on the RAPIDPoint 500 blood gas analyzer. When exhibited, the issue is usually seen during Level 1 QC /AQC, Level 1 CVM, and sometimes Level 2 CVM testing. This issue has also been observed on a small number of patient samples.
Recall Action Recall for Product Correction
Recall Action Instructions Customers are advised that the following steps can be taken to mitigate the issue:
- Install a RAPIDPoint 500 Measurement Cartridge
or
- Install a 405 Measurement Cartridge. Customers will need to arrange to get the lactate results from other instruments in their central laboratory since the Lactate parameter is not available on the RAPIDPoint 405 cartridge
or
- Use an alternative system to obtain the electrolyte values.

Since this issue is not related to the AQC cartridge, installing a new AQC cartridge will not solve the problem.
Contact Information 1800 310 300 - Siemens Technical Support Centre