Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00570-1
Product Name/Description Neoflow Flow Sensor and Neoflow Insert

Part Number: 8410179
ARTG Number: 132264
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 20/08/2014
Responsible Entity Draeger Medical Australia Pty Ltd
Reason/Issue Investigations have shown that the Neonatal Flow Sensor Insert (8410179) can leak after connecting the plug. This was caused by a tool within the Neonatal Flowsensor Insert production process.
Recall Action Recall
Recall Action Instructions Drager requested their customers to review their stock and return once replacement stock was available.

This action was undertaken prior to notifying the Therapeutic Goods Administration.
Contact Information 1800 372 437 - Drager Customer Service Australia