Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00570-1 |
Product Name/Description |
Neoflow Flow Sensor and Neoflow Insert
Part Number: 8410179 ARTG Number: 132264 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
20/08/2014 |
Responsible Entity |
|
Reason/Issue |
Investigations have shown that the Neonatal Flow Sensor Insert (8410179) can leak after connecting the plug. This was caused by a tool within the Neonatal Flowsensor Insert production process. |
Recall Action |
Recall |
Recall Action Instructions |
Drager requested their customers to review their stock and return once replacement stock was available.
This action was undertaken prior to notifying the Therapeutic Goods Administration. |
Contact Information |
1800 372 437 - Drager Customer Service Australia |