Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00566-1
Product Name/Description Tenor without scale (Mobile patient transfer lift)

All devices manufactured since 2004, not equipped with scale

ARTG Number: 199602
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 20/05/2014
Responsible Entity Huntleigh Healthcare Pty Ltd
Reason/Issue During use, the bolt in the connection between the spreader bar and the lifting arm jig is subjected to mechanical forces, which may lead to breakage. This is due to the design of the connection between spreader bar and lifting arm jig, which in long term use can cause metal fatigue in this part and the formation of micro-cracks in the thread of the bolt. This issue is only related to Tenor devices without scales.
Recall Action Recall for Product Correction
Recall Action Instructions ArjoHuntleigh will arrange for the replacement of the old spreader bar with a new spreader with scales. This action has been closed-out on 12/02/2016.
Contact Information 08 9314 4117 - ArjoHuntleigh