Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00546-1
Product Name/Description Alegria Anti-Intrinsic Factor Assay

Product Code: ORG 247-24

Lot Numbers: 24732425, 24732427 & 24733840
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 21/07/2014
Responsible Entity Immuno Pty Ltd
Reason/Issue Analysis in Quality Control has identified and confirmed that the lots listed above of our Alegria test ORG 247, anti-Intrinsic Factor, exhibit an overall lower level of optical density values. This effect bears the risk that some positive sample-values will fall below the cut-off and will falsely be read as negative.
Recall Action Recall
Recall Action Instructions Immuno requested thier customers to destroy any remaining product from the affecetd lots. Due to the risk of false negative results Immuno recommended that negative samples where clinical symptoms of pernicious anaenia should be retested. This action has been closed-out on 18/02/2016.