Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00543-1 |
Product Name/Description |
Single-width Airway Module E-miniC and Extension Modules N-FC & N-FCREC (used with CARESCAPE Monitors B850, B650 and B450, S/5 modular monitors, Engstrom Carestation, Patient Monitor B30, S/5M and FML monitors)
Single-width Airway Module E-miniC Serial Numbers: 6818561 to 6898777
Extension Modules N-FC & N-FCREC Serial Numbers: 6799191 to 6905206
ARTG Number: 166230 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
15/05/2014 |
Responsible Entity |
|
Reason/Issue |
GE Healthcare has become aware of a potential safety issue due to failure of the carbon dioxide detector associated with Single-width Airway and Extension Modules. Failure of the carbon dioxide detector in Single-width Airway and Extension modules may cause a slow continuous decrease of measured Et-/FiCO2 values. Incorrect Et CO2/Fi CO2 value may impair clinical decision making for both mechanically and spontaneously ventilated patients because of incorrect low CO2 values. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
GE is providing calibration instructions to assist their customers to determine if their module is affected. If the carbon dioxide reading is out of range after the first successful calibration procedure end users are advised to discontinue use and contact a GE service representative. GE will arrange with customers for a correction to ensure this issue does not occur. This action has been closed-out on 02/02/2016. |
Contact Information |
1800 659 465 - GE Healthcare National Call Centre |