Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00509-1
Product Name/Description Receptal System Liners and Canisters

All list numbers affected
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 8/05/2014
Responsible Entity Hospira Pty Limited
Reason/Issue Through customer complaints, Hospira has become aware of customers using the incorrect size Receptal liners with their Receptal canisters, e.g., 2L liner or 3L liner with a 1.5L canister, which has resulted in a report of one death to date. There have been no reports of death or serious injury in Australia. It is possible that if an incorrectly sized liner is used with a different size canister the lid may not fit and the vacuum/suction will not be sufficient and/or consistent. Hospira advises users to not use incorrectly sized Receptal liners with Receptal canisters. All Receptal Liner and Canister List numbers are affected by this action.
Recall Action Recall for Product Correction
Recall Action Instructions Hospira is advising hospitals to only use the correct size liners are used with the corresponding canister and to ensure that the liner is fully extended in the liner.
Contact Information 1300 046 774 - Hospira Customer Service