TGA Recall Reference RC-2014-RN-00507-1
Product Name/Description Flexi-Seal CONTROL Faecal Management System

Product code: 411107
All lot numbers

ARTG number: 163198
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 8/05/2014
Responsible Entity Convatec Australia Pty Ltd
Reason/Issue An internal assessment of product performance, including a review of customer feedback, has confirmed that Flexi-Seal CONTROL Faecal Management System is not meeting performance requirements for this product. Specifically, the Auto-Valve has not been consistently performing relative to the inflation and deflation of the device’s retention balloon. In the event that the Auto-Valve does not perform properly related to inflation and deflation of the device’s retention balloon, there is the potential that medical intervention may be necessary. In accordance with the Instructions for Use supplied with the Flexi-SealTM ControlTM Faecal Management System, the following events could occur:
- Leakage of stool around the device, and peri-anal skin breakdown;
- Rectal/anal bleeding due to pressure necrosis or ulceration of rectal or anal mucosa;
- Perforation of the bowel.
Recall Action Recall
Recall Action Instructions End users are advised to stop the use of all affected devices and return all affected products to their distributor for credit or replacement with unaffected product.