Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00502-1
Product Name/Description VITROS ALT, ALKP, AST, CHE, CK, GGT and LDH results generated on VITROS 4600 Chemistry System, VITROS 5600 Integrated System and VITROS 5,1 FS Chemistry System

VITROS 4600 Chemistry System
Product Code: 6802445
Affected Software Versions: 3.0 and lower

VITROS 5600 Integrated System
Product Code: 6802413
Affected Software Versions: 3.0 and lower

VITROS 5,1 FS Chemistry System
Product Codes: 6801375, 6801890
Affected Software Versions: 2.7 and lower

Multiple assays affected
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/05/2014
Responsible Entity Ortho-Clinical Diagnostics
Reason/Issue This recall action has been initiated by Ortho-Clinical Diagnostics, Inc. (OCD) due to the potential for biased results when a Multiple Window (MW) code is associated with results generated by the VITROSĀ® Systems listed. A MW code is a flag/error code generated when a potential issue with the results calculation is detected. This code does not suppress the sample result. A MW code is a flag/error code generated when a potential issue with the results calculation is detected. This code does not suppress the sample result.
Recall Action Recall for Product Correction
Recall Action Instructions Ortho-Clinical Diagnostics (OCD) will resolve this issue in a future software release. To assist in the interim OCD is providing a procedure for reviewing VITROS chemistry results and MW codes. This action has been closed-out on 11/02/2016.
Contact Information 1800 032 359 - Ortho-Clinical Diagnostics Customer Technical Services