Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00493-1 |
Product Name/Description |
Injectomat Agilia Range (Std, MC & TIVA), Volumat Agilia Range (Std & MC) |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
18/07/2014 |
Responsible Entity |
|
Reason/Issue |
Frensenius Kabi initiated a software upgrade due to three incident reports concerning three different issues:
Injectomat and Volumat pumps: 1. The PDMS from Picis implemented an unexpected high number of connection/disconection requests resulting in a "Watchdog" error message. 2. If the user configured the Vigilant Drug Lib software with a drug with fixed dilutions, the selection on the pump of this drug as a first step then followed by the selection of 'Drug X in mL/hr' as a second step could result in the display of a technical error.
Concerning Injectomat pumps: 3. In a high vibration environment, for example close to the warming device, the syringe installation check sensor signal can be unstablepotentially triggering a technical error (n-deg 23).
This action was undertaken prior to notifying the TGA. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Fresenius Kabi has updated the software of all affected Injectomat and Volumat devices. |
Contact Information |
02 9391 5590 - Fresenius Kabi Australia |