Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00473-1 |
Product Name/Description |
Percutaneous Sheath Introducer Kits
Batch numbers AK09903A - 23F13G0162 SI09903E – RF3064915, RF3064710
ARTG Number: 135922 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
28/04/2014 |
Responsible Entity |
|
Reason/Issue |
The manufacturer has initiated this recall due to tip damage noticed during internal inspection of the dilator tip of the Percutaneous Sheath Introducer found in the above referenced kits. Tip damage has the potential to result in vessel damage. |
Recall Action |
Recall |
Recall Action Instructions |
Mayo Healthcare is requesting end users to inspect their stock and quarantine any units from the affected batches. Mayo Healthcare will contact customers to arrange for the recovery of affected devices. |
Contact Information |
1300 360 226 - Mayo Healthcare National Customer Service |