Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00473-1
Product Name/Description Percutaneous Sheath Introducer Kits

Batch numbers
AK09903A - 23F13G0162
SI09903E – RF3064915, RF3064710

ARTG Number: 135922
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 28/04/2014
Responsible Entity Mayo Healthcare Pty Ltd
Reason/Issue The manufacturer has initiated this recall due to tip damage noticed during internal inspection of the dilator tip of the Percutaneous Sheath Introducer found in the above referenced kits. Tip damage has the potential to result in vessel damage.
Recall Action Recall
Recall Action Instructions Mayo Healthcare is requesting end users to inspect their stock and quarantine any units from the affected batches. Mayo Healthcare will contact customers to arrange for the recovery of affected devices.
Contact Information 1300 360 226 - Mayo Healthcare National Customer Service