Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00458-1
Product Name/Description Otto Bock A200 Power Wheelchair, SKIPPI and SKIPPI PLUS Power Wheelchairs

Units delivered between 20 June 2011 and 8 March 2013

ARTG Number: 177246
Recall Action Level Consumer
Recall Action Classification Class II
Recall Action Commencement Date 23/04/2014
Responsible Entity Otto Bock Australia Pty Ltd
Reason/Issue During the course of on-going market observations the front caster fork complaint rate of the three wheelchairs has increased. A detailed analysis of the reports shows that the caster forks may, in unusual circumstances, break when repeatedly driven over or against street curbs.
Recall Action Recall for Product Correction
Recall Action Instructions Ottobock Healthcare is replacing the caster fork on all affected A200 Power Wheelchair, SKIPPI and SKIPPI PLUS Power Wheelchairs. Product corrections are to be carried out by the wheelchair dealer. For more details, please see http://www.tga.gov.au/safety/alerts-device-ottobock-power-wheelchairs.htm .
Contact Information 02 8818 2817 - Otto Bock Australia Pty Ltd