Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00449-1 |
Product Name/Description |
Tension Band Pins (Internal fixation pins)
Part Numbers:
TB-1550K-S Lot Number: 223007
TB-1570K-S Lot Numbers: 220788, 230952 & 233742
TB-1590K-S Lot Numbers: 218700, 221465 & 223009
ARTG Number: 99821 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class III |
Recall Action Commencement Date |
9/05/2014 |
Responsible Entity |
|
Reason/Issue |
Specific lots of tension band pins manufactured between July 2009 and March 2011 are being recalled because they were manufactured with ASTM F138 stainless steel but should have been manufactured with ASTM A276 stainless steel. Both standards are grade 316L stainless steel and contain the exact same chemical elements but in different concentrations. Although the biocompatibility was only tested on ASTM F138 there is no evidence to indicate that ASTM A276 is not biocompatible. |
Recall Action |
Recall |
Recall Action Instructions |
Medical + Optical is contacting their customers to arrange for the return of affected devices. |
Contact Information |
1800 225 065 - Medical + Optical Instruments Australia Pty Ltd |