Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00447-1
Product Name/Description FOLFusor SV 0.5 ml/hr

Product Code: 2C4700K

All lot numbers up to and including 14B039

ARTG Number: 168893
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 17/04/2014
Responsible Entity Baxter Healthcare Pty Ltd
Reason/Issue The direction insert for the FOLFusor SV 0.5 ml/hr states that infusion rates may be up to 10% greater than the nominal infusion rate of 0.5 ml/hr. Flow rate testing indicates that infusion rates may be higher and could result in a 15% greater than nominal infusion rate.
Recall Action Recall for Product Correction
Recall Action Instructions Baxter is notifying their customers that infusion rates may be higher and could result in a 15% greater than nominal infusion rate rather than 10% as stated in the IFU. Baxter is also reiterating the factors that can contribute to over infusion. These factors along with the 15% flow rate accuracy could result in infusion rates in excess of 30% greater than the nominal flow-rate.

Baxter is currently investigating options to bring the flow rate into specification.
Contact Information 1300 789 646 - Baxter Healthcare Customer Service