Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00437-1
Product Name/Description Infant Nasal CPAP Prongs and Bubble CPAP Starter Kits

Affected model numbers / lots numbers

ARTG Numbers: 129354 & 141298
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 14/04/2014
Responsible Entity Fisher & Paykel Healthcare Pty Ltd
Reason/Issue FPH is issuing this recall notice because it has received customer reports relating to certain lot numbers of certain models of Nasal CPAP Prongs detaching from the FPH FlexiTrunk Nasal Tubing, especially in the presence of mucous and/or moisture. In the event of the Affected Prongs detaching from the Nasal Tubing, therapy is likely to be interrupted. This may lead to a patient becoming hypoxic.
Recall Action Recall
Recall Action Instructions FPH is requesting their customers to identify any units form the affected lot and destroy them. An FPH Representative can assist in sourcing a replacement product.
Contact Information 03 9879 5022 - Fisher & Paykel Healthcare