Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00436-2
Product Name/Description DigitalDiagnost (Digital X-ray System)

ARTG number: 117662
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 11/04/2014
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue The footboard of the Patient support for stitching can be fixed in upper position. When the fix clamp (hook) is not correctly adjusted into its correct position, the footboard is not securely held. If it falls down, it may hurt the operator's or patient's foot (e.g. when moving the stand or when the operator fails to fix it correctly upon moving the board up).
Recall Action Recall for Product Correction
Recall Action Instructions Philips is advising their customers to ensure that the hook is securely holding the footplate when in a vertical position. A redesigned hook is under development and will be fitted when complete. This action has been closed-out on 18/02/2016.
Contact Information 1800 251 400 - Philips Customer Care Centre