Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00414-1
Product Name/Description enGen Laboratory Automation Systems configured with Data Innovation, LLC Instrument Manager (IM) v8.06

Product Code: 6802883

Lot Numbers: enGen Select Versions 3.2.2 and below & enGen Custom versions 3.2.2 and below
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/04/2014
Responsible Entity Ortho-Clinical Diagnostics
Reason/Issue The enGen Laboratory Automation System is an automated track which integrates clinical analysers and the customer’s LIS (Laboratory Information System). It centrally controls pre-analytical and post-analytical processing as well as sample transport to multiple independent analysers and devices. Ortho Clinical Diagnostics (OCD) internal testing has determined that when a sample requires follow-up action, such as dilution or retesting due to a mechanical issue, it is possible for that sample to be routed to the sample storage drawer (Storage) prior to the completion of the necessary pending action. A delay in patient test result(s) reporting could occur.
Recall Action Recall for Product Correction
Recall Action Instructions Laboratories are advised to manually rerun samples that have been prematurely stored. A software modification is currently in development and will be provided once complete. This action has been closed-out on 11/02/2016.