Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00413-1 |
Product Name/Description |
VITROS Chemistry Products Specialty Diluent and VITROS Chemistry Products FS Diluent Pack 3
VITROS Chemistry Products Specialty Diluent Product Code: 8559825 Lot Number: F3168 Expiry Date: 30 Sep 2014
VITROS Chemistry Products FS Diluent Pack 3 Product Code: 6801754 Lot Number: 01-3266 Expiry Date: 23 Sep 2014 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
7/04/2014 |
Responsible Entity |
|
Reason/Issue |
Ortho Clinical Diagnostics Inc confirmed reports of lower than expected C-Reactive Protein (CRP) results when using the affected lots of VITROS Specialty Diluent and VITROS FS Diluent Pack 3 to dilute samples tested on VITROS CRP Slides. Internal testing determined that results using VITROS CRP Slides may be negatively biased by approximately 51% using a 3X dilution factor. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Ortho Clinical Diagnostics is requesting their customers to discontinue use and discard any remaining stock from the affected lots of diluent. Replacements from unaffected lots are available. This action has been closed-out on 11/04/2016 |
Contact Information |
1800 032 359 - Ortho Clinical Diagnostics Customer Technical Services |