Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00406-1
Product Name/Description Monitor AC2000 Spring Arm

Part Numbers: G19051 and G19262
All serial numbers affected

Part Number: G93358
All serial numbers until 0111 066 23639 affected

Part Number: G93359
All serial numbers until 0111 056 21343 affected

ARTG Number: 132260
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 7/04/2014
Responsible Entity Draeger Medical Australia Pty Ltd
Reason/Issue Draeger Medical Australia has been informed of three cases where a welded seam in the spring-loaded arm of the monitor holder broke resulting in the monitor falling. Although no injury has been reported, the risk of injury to persons positioned beneath the monitor cannot be excluded.
Recall Action Recall
Recall Action Instructions A Drager service representative will inspect all potentially affected spring-loaded arms. A replacement arm will be provided for all devices determined to be affected.
Contact Information 1800 800 327 - Draeger Medical Australia Pty Ltd