Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00405-1 |
Product Name/Description |
ADAC Vertex Classic, Vertex Plus, Vertex V60 and Solus Systems
ARTG Number: 117642 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
1/04/2014 |
Responsible Entity |
|
Reason/Issue |
Philips Healthcare received a report from the field about an ADAC Vertex Plus system. During clinical use in the relative 180 degree configuration of the detector heads, the radius drive belt in the superior positioned head slipped off the idler pulley, allowing the detector head to drift down slowly (approximately 1 cm per minute) towards the patient. This resulted in the operator having to perform an emergency removal of the patient from the system. There have been no reports of serious injury or death as a result of this situation. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Hospitals are requested to cease use of "Relative 180 degree SPECT" and "pinhole collimator" procedures until the implementation of the correction. Philips will replace the two radius drive belts and idler pulleys to correct the problem. |
Contact Information |
1800 251 400 - Philips Customer Care Centre |