Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00389-1
Product Name/Description OEC 9900 Workstation Monitor Spring Arm Screws

All OEC 9900 systems manufactured between October 9, 2013 and March 20, 2014, and all OEC 9900 systems that had an extension arm replaced between October 9, 2013 and March 20, 2014 are affected.

ARTG Number: 132698
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 2/04/2014
Responsible Entity GE Healthcare Australia Pty Ltd
Reason/Issue The screws that attach the workstation monitor spring arm to the extension arm may fail, causing the spring arm and monitor to fall forward.
Recall Action Recall for Product Correction
Recall Action Instructions A GE HealthCare Field Engineer will replace the screwS to ensure that the system is operating safely and meets all specifications. Until the correction is complete, end users are advise to use caution when positioning or extending the arm. In addition, refrain from positioning staff or patients directly underneath or in front of a work station monitor.
Contact Information 1800 659 465 - GE Healthcare National Call Centre