Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00383-1
Product Name/Description ARTIS Zeego (Fluoroscopic angiographic x-ray system)

Serial number of the motor connectors lower than 946925
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 1/04/2014
Responsible Entity Siemens Ltd
Reason/Issue Siemens Healthcare has determined that the crimping of motor connectors A1 to A3 may have been incorrectly carried out. Incorrectly executed crimping where the crimping tool was used on the wrong point of the crimp contact has led to substandard contacting. The resulting increase in interference resistance can cause heat build up and potentially lead to scorching of the corresponding cables.
Recall Action Recall for Product Correction
Recall Action Instructions Siemens will be inspecting the connectors to ensure they are correctly crimped at the A1 to A3 motor terminals. A Siemens representative will be in touch to schedule the visit.
Contact Information 1800 310 300 - Siemens Technical Support