Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00374-1 |
Product Name/Description |
Alere Triage BNP Test for Beckman Coulter (intended to measure B-type Natriuretic Peptide levels)
Catalogue Number: BS98200
Lot Number: 331265
ARTG Number: 215842 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
31/03/2014 |
Responsible Entity |
|
Reason/Issue |
The affected lot may not recover within range when tested using certain commercially available controls. For example, this lot may have lower than expected results when using Bio-Rad's Liquichek Cardiac Marker Plus and Liquichek Cardiac Marker Plus LT controls. This lot may also have higher than expected results when using Thermo Scientific MAS Cardiolmmune XL Controls. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Alere is alerting their customers to the issue and advising that the affected lots should only be used with Alere Triage BNP QC Controls. It is also recommended that all borderline control and patient results are repeated using an alternative method. |
Contact Information |
07 3363 7711 - Alere Technical Support |