TGA Recall Reference RC-2014-RN-00364-1
Product Name/Description IntelliSpace Portal with AutoSPECT Pro reconstruction application (image viewing system for diagnostic imaging, x-ray and computed tomography)

Affected IntelliSpace Portal software versions: v5.0.0, v5.0.1, v5.0.2 and v6.0.0

ARTG Number: 188684
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 26/03/2014
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal with the software versions mentioned above. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on IntelliSpace Portal unless the data is acquired with VXGP collimators. The occurrence of either problem may result in image quality degradation that could result in misdiagnosis under certain conditions.
Recall Action Recall for Product Correction
Recall Action Instructions Philips is providing work around instructions to ensure the continued safe use of the AutoSPECT Pro until a software update can be provided to correct the issue. This action has been closed-out on 12/02/2016.