Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00361-1 |
Product Name/Description |
Shape Arm Kit Left and Shape Arm Kit Right (Mounting/positioning system for flat panel monitors)
Shape Arm Kit Left Catalogue Number: 0682000976 Serial Numbers: 080924DC001 through 120227DC025
Shape Arm Kit Right Catalogue Number: 0682000977 Serial Numbers: 080909DB001 through 131021DB013
ARTG Number: 118878 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
25/03/2014 |
Responsible Entity |
|
Reason/Issue |
Stryker has become aware that there is a remote possibility that the lock washer and screw that secures the monitor onto the SHAPE Arm may not be present. If the SHAPE Arm lock washer and screw are not present the monitor could separate and dislodge from the Arm potentially resulting in injury. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Stryker is inspecting all affected Shape Arms to determine if the lock washer and screw are missing. If missing a new lock washer and screw will be fitted. |
Contact Information |
1800 803 601 - Stryker Regulatory Affairs & Quality Assurance |