Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00295-1
Product Name/Description Endo-Scrub2 Sheaths and Tubing Set

Sheath Models #’s (w/ tubing): 1912000, 1912004, 1912008, 1912032, 1912033, 1912038, 1912040, 1912050

Tubing Model #’s: 1912030

Multiple lot numbers

ARTG numbers: 125734 and126694
Recall Action Level Hospital
Recall Action Classification Class III
Recall Action Commencement Date 18/03/2014
Responsible Entity Medtronic Australasia Pty Ltd
Reason/Issue The luer connector on the end of the tubing that attaches to the sheath may break off, either prior to or during use. Once the connector breaks, the flow of saline to the sheath either ceases or is reduced and leakage occurs at the fracture site.
Recall Action Recall for Product Correction
Recall Action Instructions Medtronic is requesting end users to inspect the luer connection for breakage at the point of use. Any broken tubing sets are to be discarded. New tubing sets will be provided to replace broken devices.
Contact Information 02 9857 9234 - Medtronic Australasia Pty Ltd