Type of Product |
Medical Device |
TGA Recall Reference |
RC-2014-RN-00287-1 |
Product Name/Description |
syngo Dynamics software version v9.5 using Sensis Integration (Radiology picture archiving and communication system)
ARTG Number: 120332 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
7/03/2014 |
Responsible Entity |
|
Reason/Issue |
A software error could allow for two different patients to be displayed in syngo Dynamics where, images shown for one patient and reporting system data (worksheet and/or report) shown for a different patient potentially resulting in misdiagnosis due to mix-up in patient data. When this error occurs, a previously opened study is not properly disconnected from the Sensis hemodynamics system. Only customers using Sensis-syngo Dynamics integration will face the occasional probability of this error occurring. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
demographics to review the image toolbar and reporting system, and to ensure that toolbar "auto-hide" is disabled. If a patient mismatch is observed, users are advised to close and reopen the study. A software update will be implemented as a corrective measure. This action has been closed-out on 29/01/2016. |
Contact Information |
1800 310 300 - Siemens Technical Support Centre |