Product Name/Description |
Indiko and Indiko Plus Chemistry Analysers |
Recall Action Level |
Hospital |
Recall Action Classification |
Class I |
Recall Action Commencement Date |
26/02/2014 |
Responsible Entity |
|
Reason/Issue |
Internal investigations have revealed that cuvettes intermittently may be used twice, and as a consequence there is the potential for erroneous results to be generated without the operator being aware that this has occurred. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Thermofisher is advising all end users to immediately stop using any analyser with software versions prior to 5.1.1. Thermo Fisher will provide their customers with a mandatory update to software version 5.1.1. Technical Support will be contacting customers as a matter of urgency to schedule the update. |