Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00212-1
Product Name/Description Indiko and Indiko Plus Chemistry Analysers
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 26/02/2014
Responsible Entity Thermo Fisher Scientific Australia Pty Ltd
Reason/Issue Internal investigations have revealed that cuvettes intermittently may be used twice, and as a consequence there is the potential for erroneous results to be generated without the operator being aware that this has occurred.
Recall Action Recall for Product Correction
Recall Action Instructions Thermofisher is advising all end users to immediately stop using any analyser with software versions prior to 5.1.1. Thermo Fisher will provide their customers with a mandatory update to software version 5.1.1. Technical Support will be contacting customers as a matter of urgency to schedule the update.
Contact Information 03 9757 4300 - Thermo Fisher Scientific Australia