Type of Product Medical Device
TGA Recall Reference RC-2014-RN-00209-1
Product Name/Description Philips HeartStart XL, Defibrillator/Monitor

Model Number: M4735A

Units Affected: Units manufactured by Philips between May, 2013 and November, 2013

Serial numbers within the range: US00101665 to US00608704
Recall Action Level Hospital
Recall Action Classification Class I
Recall Action Commencement Date 25/02/2014
Responsible Entity Philips Electronics Australia Ltd
Reason/Issue An internal component on the HeartStart XL power board may malfunction potentially affecting the ability to deliver therapy. Specifically, the XL may be unable to charge and deliver a shock. In addition, when used for external pacing, the malfunction may result in a loss of capture.
Recall Action Recall for Product Correction
Recall Action Instructions Philips will be installing replacement power boards to correct the problem. A Philips representative will be in contact with customers to make arrangements for the correction. While awaiting the replacement power board for the device, users may continue to use the HeartStart XL. Identify a readily available backup defibrillator to use in the event the affected HeartStart XL fails to deliver therapy. This action has been closed-out on 11/02/2016.
Contact Information 1800 251 400 - Phillips Customer Care Centre